A Multiple Local Models Approach to Accuracy Improvement in Continuous Glucose Monitoring

2011

Continuous glucose monitoring (CGM) devices estimate plasma glucose (PG) from measurements in compartments alternative to blood. The accuracy of currently available CGM is yet unsatisfactory and may depend on the implemented calibration algorithms, which do not compensate adequately for the differences of glucose dynamics between the compartments. Here we propose and validate an innovative calibration algorithm for the improvement of CGM performance.CGM data from GlucoDay(®) (A. Menarini, Florence, Italy) and paired reference PG have been obtained from eight subjects without diabetes during eu-, hypo-, and hyperglycemic hyperinsulinemic clamps. A calibration algorithm based on a dynamic glo…

Blood Glucosemedicine.medical_specialtyCalibration (statistics)Endocrinology Diabetes and MetabolismMonitoring Ambulatory030209 endocrinology & metabolismBiosensing TechniquesAccuracy improvementSensitivity and Specificity01 natural sciencesGlobal model03 medical and health sciences0302 clinical medicineEndocrinologyInternal medicineBlood Glucose Self-MonitoringDiabetes MellitusmedicineHumansGlucose dynamicsPlasma glucoseContinuous glucose monitoringbusiness.industryBlood Glucose Self-Monitoring010401 analytical chemistryReproducibility of ResultsPattern recognition0104 chemical sciencesMedical Laboratory TechnologyEndocrinologyCalibration algorithmArtificial intelligencebusinessAlgorithmsDiabetes Technology & Therapeutics